Evaluation of transport ventilators at mild simulated altitude: a bench study in a hypobaric chamber

Respir Care. 2014 Aug;59(8):1233-41. doi: 10.4187/respcare.02985.

Abstract

Background: Previous studies on ventilators used for air transport showed significant effects of altitude, in particular with regard to accuracy of the tidal volume (VT) and breathing frequency. The aim of the study was to evaluate transport ventilators under hypobaric conditions.

Methods: We conducted a bench study of 6 transport ventilators in a Comex hypobaric chamber to simulate mild altitude (1,500 m [4,920 feet] and 2,500 m [8,200 feet]). The ventilators were connected to a test lung to evaluate their accuracy: (1) to deliver a set VT under normal resistance and compliance conditions at F(IO2) = 0.6 and 1, (2) to establish a set PEEP (0, 5, 10, and 15 cm H2O), and (3) to establish a set inspiratory pressure in pressure controlled mode, (4) at a F(IO2) setting, and (5) and at a frequency setting.

Results: Four ventilators kept an average relative error in VT of < 10% without effect of altitude. The Medumat ventilator was affected by the altitude only at F(IO2) = 1. The Osiris 3 ventilator had > 40% error even at 1,500 m. We found no change in frequency as a function of altitude for any ventilators studied. No clinically important differences were found between all altitudes with the PEEP or inspiratory pressure setting. Although F(IO2) was affected by altitude, the average error did not exceed 11%, and it is unclear whether this fact is an experimental artifact.

Conclusions: We have shown that most of the new transport ventilators tested require no setting adjustment at moderate altitude and are as safe at altitude as at sea level under normal respiratory conditions. Older technologies still deliver more volume with altitude in volumetric mode.

Publication types

  • Comparative Study

MeSH terms

  • Air Ambulances*
  • Altitude*
  • Atmosphere Exposure Chambers
  • Atmospheric Pressure
  • Equipment Failure Analysis
  • Humans
  • Models, Biological
  • Respiratory Rate
  • Tidal Volume
  • Ventilators, Mechanical*