Multicenter validation of the VITEK MS v2.0 MALDI-TOF mass spectrometry system for the identification of fastidious gram-negative bacteria

Diagn Microbiol Infect Dis. 2014 Feb;78(2):129-31. doi: 10.1016/j.diagmicrobio.2013.08.013. Epub 2013 Dec 8.

Abstract

The VITEK MS v2.0 MALDI-TOF mass spectrometry system's performance in identifying fastidious gram-negative bacteria was evaluated in a multicenter study. Compared with the reference method (DNA sequencing), the VITEK MS system provided an accurate, species-level identification for 96% of 226 isolates; an additional 1% were accurately identified to the genus level.

Keywords: Fastidious; Gram-negative bacilli; Identification; MALDI-TOF MS; VITEK.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Bacterial Typing Techniques / methods*
  • Bacterial Typing Techniques / standards
  • Gram-Negative Bacteria / classification*
  • Gram-Negative Bacteria / genetics
  • Gram-Negative Bacteria / isolation & purification
  • RNA, Bacterial
  • RNA, Ribosomal, 16S
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / standards