Determination of thiomersal by flow injection coupled with microwave-assisted photochemical online oxidative decomposition of organic mercury and cold vapor atomic fluorescence spectroscopy

Anal Chim Acta. 2013 Dec 4:804:66-9. doi: 10.1016/j.aca.2013.10.018. Epub 2013 Oct 18.

Abstract

We developed a flow injection (FI) method for the determination of thiomersal (sodium ethylmercurithiosalicylate, C9H9HgNaO2S) based on the UV/microwave (MW) photochemical, online oxidation of organic mercury, followed by cold vapor generation atomic fluorescence spectrometry (CVG-AFS) detection. Thiomersal was quantitatively converted in the MW/UV process to Hg(II), with a yield of 97±3%. This reaction was followed by the reduction of Hg(II) to Hg(0) performed in a knotted reaction coil with NaBH4 solution, and AFS detection in an Ar/H2 miniaturized flame. The method was linear in the 0.01-2 μg mL(-1) range, with a LOD of 0.003 μg mL(-1). This method has been applied to the determination of thiomersal in ophthalmic solutions, with recoveries ranging between 97% and 101%. We found a mercury concentration in commercial ophthalmic solutions ranging between 7.5 and 59.0 μg mL(-1).

Keywords: Atomic fluorescence spectrometry; Flow injection analysis; Mercury oxidation; Organic mercury-compounds; Thiomersal.

MeSH terms

  • Chromatography, High Pressure Liquid
  • Flow Injection Analysis
  • Mercury / analysis*
  • Microwaves*
  • Oxidation-Reduction
  • Spectrometry, Fluorescence / methods*
  • Thimerosal / analysis*

Substances

  • Thimerosal
  • Mercury