In vivo evaluation of immediately loaded stainless steel and titanium orthodontic screws in a growing bone

PLoS One. 2013 Oct 4;8(10):e76223. doi: 10.1371/journal.pone.0076223. eCollection 2013.

Abstract

The present work intends to evaluate the use of immediate loaded orthodontic screws in a growing model, and to study the specific bone response. Thirty-two screws (half of stainless steel and half of titanium) were inserted in the alveolar bone of 8 growing pigs. The devices were immediately loaded with a 100 g orthodontic force. Two loading periods were assessed: 4 and 12 weeks. Both systems of screws were clinically assessed. Histological observations and histomorphometric analysis evaluated the percent of "bone-to-implant contact" and static and dynamic bone parameters in the vicinity of the devices (test zone) and in a bone area located 1.5 cm posterior to the devices (control zone). Both systems exhibit similar responses for the survival rate; 87.5% and 81.3% for stainless steel and titanium respectively (p = 0.64; 4-week period), and 62.5% and 50.0% for stainless steel and titanium respectively (p = 0.09; 12-week period). No significant differences between the devices were found regarding the percent of "bone-to-implant contact" (p = 0.1) or the static and dynamic bone parameters. However, the 5% threshold of "bone-to-implant contact" was obtained after 4 weeks with the stainless steel devices, leading to increased survival rate values. Bone in the vicinity of the miniscrew implants showed evidence of a significant increase in bone trabecular thickness when compared to bone in the control zone (p = 0.05). In our study, it is likely that increased trabecular thickness is a way for low density bone to respond to the stress induced by loading.

MeSH terms

  • Animals
  • Bone Screws*
  • Stainless Steel / chemistry*
  • Swine
  • Titanium / chemistry*

Substances

  • Stainless Steel
  • Titanium

Grants and funding

The authors have no support or funding to report.