Determination of asymmetric and symmetric dimethylarginines in human plasma by HPLC with electrochemical detection

J Sep Sci. 2013 Dec;36(23):3696-701. doi: 10.1002/jssc.201300813. Epub 2013 Oct 20.

Abstract

A new HPLC method was developed and validated for the determination of asymmetric and symmetric dimethylarginines and l-arginine in human plasma. After SPE and evaporation of the eluate, the samples were derivatised with an o-phthaldialdehyde reagent containing 3-mercaptopropionic acid. The derivatives formed were analysed by isocratic RP-HPLC with electrochemical detection at +320 mV. The mobile phase consisted of 50 mM phosphate buffer (pH 6.1) containing 10% v/v acetonitrile, the flow rate was 1 mL/min. The retention times of all compounds including monomethylarginine (internal standard) were <24 min. The LODs (S/N 3:1) were 0.012 μM for both dimethylarginines and 0.013 μM for L-arginine; the linearity of the method was from 0.1 to 20 μM for both dimethylarginines and from 1 to 200 μM for L-arginine. Absolute extraction recoveries measured for all analytes ranged from 85 to 88%.

Keywords: Arginine; Asymmetric dimethylarginine; Electrochemical detection; HPLC; Symmetric dimethylarginine.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Arginine / analogs & derivatives*
  • Arginine / blood
  • Blood Chemical Analysis / instrumentation*
  • Blood Chemical Analysis / methods*
  • Chromatography, High Pressure Liquid*
  • Electrochemical Techniques*
  • Humans
  • Limit of Detection
  • Reference Standards