Excision of a permanent inferior vena cava filter with multiple vena caval perforations

Ann Vasc Surg. 2014 Jan;28(1):261.e7-9. doi: 10.1016/j.avsg.2013.05.004. Epub 2013 Sep 29.

Abstract

Perforation of an inferior vena cava filter by one the filter device hooks is a recognized possible complication of this device. We describe a case of surgical excision of a permanent inferior vena cava filter associated with multiple perforations of surrounding structures by each of the 6 hooks of the device. Structures affected include the third lumbar vertebral body, transverse mesocolon, the infrarenal aorta, the duodenum, and the psoas muscle. A thorough understanding of the filter design and adequate preoperative imaging were vital in planning the safe surgical excision of this device.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Aorta / injuries
  • Colon / injuries
  • Device Removal*
  • Duodenum / injuries
  • Female
  • Humans
  • Intestinal Perforation / etiology
  • Lumbar Vertebrae / injuries
  • Prosthesis Design
  • Psoas Muscles / injuries
  • Tomography, X-Ray Computed
  • Treatment Outcome
  • Vascular System Injuries / diagnosis
  • Vascular System Injuries / etiology
  • Vascular System Injuries / surgery*
  • Vena Cava Filters / adverse effects*
  • Vena Cava, Inferior / diagnostic imaging
  • Vena Cava, Inferior / injuries
  • Vena Cava, Inferior / surgery*