Comparison of the accuracy for three dental impression techniques and index: an in vitro study

J Prosthodont Res. 2013 Oct;57(4):268-74. doi: 10.1016/j.jpor.2013.07.001. Epub 2013 Sep 26.

Abstract

Objectives: This in vitro study compared the dimensional accuracy of stone index (I) and three impression techniques: tapered impression copings (T), squared impression copings (S) and modified squared impression copings (MS) for implant-supported prostheses.

Methods: A master cast, with four parallel implant abutment analogs and a passive framework, were fabricated. Vinyl polysiloxane impression material was used for all impressions with two metal stock trays (open and closed tray). Four groups (I, T, S and MS) were tested (n=5). A metallic framework was seated on each of the casts, one abutment screw was tightened, and the gap between the analog of implant and the framework was measured with a stereomicroscope. The groups' measurements (80 gap values) were analyzed using software (LeicaQWin - Leica Imaging Systems Ltd.) that received the images of a video camera coupled to a Leica stereomicroscope at 100× magnification. The results were statistically analyzed with Kruskal-Wallis One Way ANOVA on Ranks test followed by Dunn's Method, 0.05.

Results: The mean values of abutment/framework interface gaps were: Master Cast=32 μm (SD 2); Group I=45 μm (SD 3); Group T=78 μm (SD 25); Group S=134 μm (SD 30); Group MS=143 μm (SD 27). No significant difference was detected among Index and Master Cast (P=.05).

Conclusion: Under the limitations of this study, it could be suggested that a more accurate working cast is possible using tapered impression copings techniques and stone index.

Keywords: Dental implant; Impression material; Impression technique; Transfer coping.

Publication types

  • Comparative Study

MeSH terms

  • Dental Casting Investment
  • Dental Impression Materials
  • Dental Impression Technique*

Substances

  • Dental Casting Investment
  • Dental Impression Materials