Analytical optimization of the ablation efficiency at normal and non-normal incidence for generic super Gaussian beam profiles

Biomed Opt Express. 2013 Jul 24;4(8):1422-33. doi: 10.1364/BOE.4.001422. eCollection 2013.

Abstract

We suggest a general method to determine the optimum laser parameters for maximizing the ablation efficiency for different materials (in particular human cornea) at different incidence angles. The model is comprehensive and incorporates laser beam characteristics and ablative spot properties. The model further provides a method to convert energy fluctuations during ablation to equivalent ablation deviations in the cornea. The proposed model can be used for calibration, verification and validation purposes of laser systems used for ablation processes at relatively low cost and would directly improve the quality of results.

Keywords: (140.3295) Laser beam characterization; (140.3425) Laser stabilization; (170.1020) Ablation of tissue; (330.7335) Visual optics, refractive surgery.