Comparison of regenerated and non-regenerated oxidized cellulose hemostatic agents

Eur Surg. 2013;45(4):213-220. doi: 10.1007/s10353-013-0222-z. Epub 2013 Jul 4.

Abstract

Background: Oxidized cellulose is a well known and widely used surgical hemostat. It is available in many forms, but manufactured using either a nonregenerated or regenerated process.

Objective: This study compares the fiber structure, pH in solution, bactericidal effectiveness, and hemostatic effectiveness of an oxidized nonregenerated cellulose (ONRC; Traumastem®) and an oxidized regenerated cellulose (ORC; Surgicel® Original).

Methods: In vitro, fiber structures were compared using scanning electron microscopy, pH of phosphate buffer solution (PBS) and human plasma were measured after each cellulose was submerged, and bactericidal effect was measured by plating each cellulose with four bacteria. In vivo, time to hemostasis and hemostatic success were compared using a general surgery nonheparinized porcine liver abrasion model and a peripheral vascular surgery heparinized leporine femoral vessel bleeding model.

Results: Ultrastructure of ONRC fiber is frayed, while ORC is smooth. ORC pH is statistically more acidic than ONRC in PBS, but equal in plasma. No difference in bactericidal effectiveness was observed. In vivo, ONRC provided superior time to hemostasis relative to ORC (211.2 vs 384.6 s, N = 60/group) in the general surgery model; and superior hemostatic success relative to ORC at 30 (60 vs. 15 %; OR: 13.5; 95 % CI: 3.72-49.1, N = 40/group), 60 (85 vs. 37.5 %; OR: 12.3; 95 % CI: 3.66-41.6), and 90 s (97.5 vs 70.0 %; OR: 21.1, 95 % CI: 2.28-195.9) in the peripheral vascular model.

Conclusion: ONRC provides superior hemostasis and equivalent bactericidal effectiveness relative to ORC, which is likely due to its fiber structure than acidity.

Keywords: Cellulose; Celstat; Fibrin pad; Hemostasis; Liver abrasion; Liver square; Oxidized cellulose; Surgicel; Traumastem.