The 4-in-5 mother-child technique: 5 Fr transradial coronary intervention for complex lesions using a 4 Fr child catheter

J Invasive Cardiol. 2013 Aug;25(8):406-8.

Abstract

Small 5 Fr guiding catheters are now used in percutaneous coronary intervention (PCI). However, when treating severely calcified and/or tortuous complex coronary lesions, the back-up support of a 5 Fr guiding catheter is often insufficient. We previously developed a 4 Fr child catheter (Kiwami ST-01; Terumo Corporation) and proposed a 4 Fr double-coaxial technique ("mother-child" configuration) to facilitate stent implantation in treating complex coronary lesions. In this manuscript, we report 2 cases of 5 Fr transradial coronary intervention in which stent delivery failed with conventional techniques. In these 2 cases, our 4 Fr double-coaxial technique was employed, and stents were successfully delivered to the target lesions. To the best of our knowledge, this is the first report demonstrating the potential of the 4-in-5 technique during 5 Fr PCI.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cardiac Catheters / classification*
  • Coronary Angiography
  • Coronary Stenosis / therapy*
  • Equipment Design
  • Female
  • Humans
  • Male
  • Middle Aged
  • Percutaneous Coronary Intervention / instrumentation*
  • Percutaneous Coronary Intervention / methods
  • Radial Artery
  • Stents*
  • Treatment Outcome