[Design of high-efficiency double compound parabolic concentrator system in near infrared noninvasive biochemical analysis]

Guang Pu Xue Yu Guang Pu Fen Xi. 2013 May;33(5):1406-10.
[Article in Chinese]

Abstract

High signal-to-noise ratio (SNR) of system is necessary to obtain accurate blood components in near infrared noninvasive biochemical analysis. In order to improve SNR of analytical system, high-efficiency double compound parabolic concentrator (DCPC) system was researched, which was aimed at increasing light utilization efficiency. Firstly, with the request of collection efficiency in near infrared noninvasive biochemical analysis, the characteristic of emergent rays through compound parabolic concentrator (CPC) was analyzed. Then the maximum focusing angle range of the first stage CPC was determined. Secondly, the light utilization efficiency of truncated type was compared with standard DCPC, thus the best structure parameters of DCPC system were optimized. Lastly, combined with optical parameters of skin tissue, calculations were operated when incident wavelength is 1 000 nm. The light utilization efficiency of DCPC system, CPC-focusing mirror system, and non-optical collecting system was calculated. The results show that the light utilization efficiency of the three optical systems is 1.46%, 0.84% and 0.26% respectively. So DCPC system enhances collecting ability for human diffuse reflection light, and helps improve SNR of noninvasive biochemical analysis system and overall analysis accuracy effectively.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Chemical Analysis / instrumentation*
  • Blood Chemical Analysis / methods*
  • Calibration
  • Equipment Design
  • Humans
  • Optical Phenomena
  • Skin / blood supply
  • Spectroscopy, Near-Infrared / methods*