An overview of wideband immittance measurements techniques and terminology: you say absorbance, I say reflectance

Ear Hear. 2013 Jul;34 Suppl 1(0 1):9S-16S. doi: 10.1097/AUD.0b013e31829d5a14.

Abstract

This article reviews the relationships among different acoustic measurements of the mobility of the tympanic membrane, including impedance, admittance, reflectance, and absorbance, which the authors group under the rubric of immittance measures. Each of these quantities is defined and related to the others. The relationship is most easily grasped in terms of a straight rigid ear canal of uniform area terminated by a uniform middle ear immittance placed perpendicular to the long axis of the ear canal. Complications due to variations from this geometry are discussed. Different methods for measuring these quantities are described, and the assumptions inherent within each method are made explicit. The benefits of wideband measurements of these quantities are described, as are the benefits and limitations of different components of immittance and reflectance/absorbance. While power reflectance (the square of the magnitude of pressure reflectance) is relatively invariant along the length of the ear canal, it has the disadvantage that it ignores phase information that may be useful in assessing the presence of acoustic leaks in ear-canal measurements and identifying other potential error sources. A combination of reflectance and impedance magnitude and angle give a more complete description of the middle ear from measurements in the ear canal.

Publication types

  • Review

MeSH terms

  • Acoustic Impedance Tests / methods*
  • Audiology / instrumentation*
  • Hearing Disorders / diagnosis*
  • Humans
  • Terminology as Topic*