Impact of electrocardiogram-gated multi-slice computed tomography-based aortic annular measurement in the evaluation of paravalvular leakage following transcatheter aortic valve replacement: the efficacy of the OverSized AortiC Annular ratio (OSACA ratio) in TAVR

J Card Surg. 2013 Jul;28(4):373-9. doi: 10.1111/jocs.12143.

Abstract

Background: Even mild paravalvular leakage (PVL) after transcatheter aortic valve replacement (TAVR) is associated with increased late mortality. Electrocardiogram-gated multi-slice computed tomography (MSCT) enables detailed aortic annulus assessment. We describe the impact of MSCT for PVL following TAVR.

Methods: Congruence between the prosthesis and annulus diameters affects PVL; therefore, we calculated the OverSized AortiC Annular ratio (OSACA ratio) and OSACA (transesophageal echocardiography, TEE) ratio as prosthesis diameter/annulus diameter on MSCT or TEE, respectively, and compared their relationship with PVL ≤ trace following TAVR.

Results: Of 36 consecutive patients undergoing TAVR (Group A), the occurrence of PVL ≤ trace (33.3%) was significantly related to the OSACA ratio (p = 0.00020). In receiver-operating characteristics analysis, the cutoff value of 1.03 for the OSACA ratio had the highest sum of sensitivity (75.0%) and specificity (91.7%; AUC = 0.87) with significantly higher discriminatory performance for PVL as compared to the OSACA (TEE) ratio (AUC = 0.69, p = 0.028). In nine consecutive patients (Group B) undergoing TAVR based on guidelines formulated from our experience with Group A, PVL ≤ trace was significantly more frequent (88.9%) than that in Group A (p = 0.0060).

Conclusions: The OSACA ratio has a significantly higher discriminatory performance for PVL ≤ trace than the OSACA (TEE) ratio, and aortic annular measurement from MSCT is more accurate than that from TEE.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging*
  • Aortic Valve / pathology*
  • Echocardiography, Transesophageal
  • Electrocardiography / methods*
  • Female
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Multidetector Computed Tomography / methods*
  • Postoperative Complications*
  • Prosthesis Design / adverse effects*
  • Prosthesis Design / methods*
  • Prosthesis Failure / etiology*
  • Prosthesis Fitting / adverse effects*
  • Prosthesis Fitting / methods*
  • Retrospective Studies