Single-institution, 22-year follow-up of 786 CarboMedics mechanical valves used for both primary surgery and reoperation

J Thorac Cardiovasc Surg. 2014 May;147(5):1493-8. doi: 10.1016/j.jtcvs.2013.05.017. Epub 2013 Jul 13.

Abstract

Objectives: The long-term (>20 years) results for CarboMedics mechanical valves (Sorin Group, Milano, Italy) used for both primary surgery and reoperation have never been reported or compared.

Methods: Since 1990, a total of 787 CarboMedics valves have been implanted in 694 patients for aortic valve replacement, including 19 redo cases in 220 patients; for mitral valve replacement, including 108 redo cases in 381 patents; and for double (aortic and mitral) valve replacement, including 29 redo cases in 93 patients. The follow-up data were complete for 7201 patient-years in 99.3% of the patients.

Results: The hospital mortality rate of the aortic, mitral, and double valve replacement groups was 0.9%, 3.7%, and 4.3%, respectively. The corresponding freedom from valve-related morbidity rates in each group were 66.0%, 40.6%, and 48.0% at 20 years (P = .0206). A higher incidence of paravalvular leakage was observed in the mitral and double valve replacement groups than in the aortic valve replacement group (P = .0019). Of the cases of mitral paravalvular leakage after single mitral valve replacement, 97% occurred after redo single mitral valve replacement; 73% of the cases of mitral paravalvular leakage after double valve replacement occurred after redo double valve replacement.

Conclusions: CarboMedics mechanical valves used for both primary surgery and reoperation for aortic, mitral, and double valve replacement can achieve satisfactory early and long-term results, even 20 years after surgery. Care should be taken, however, to prevent paravalvular leakage in the mitral position during reoperation.

MeSH terms

  • Aortic Valve / surgery*
  • Female
  • Heart Valve Diseases / diagnosis
  • Heart Valve Diseases / mortality
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hospital Mortality
  • Humans
  • Japan
  • Male
  • Middle Aged
  • Mitral Valve / surgery*
  • Postoperative Complications / diagnosis
  • Postoperative Complications / mortality
  • Postoperative Complications / surgery*
  • Prosthesis Design
  • Reoperation
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome