Combined solid-phase microextraction and high-performance liquid chromatography with ultroviolet detection for simultaneous analysis of clenbuterol, salbutamol and ractopamine in pig samples

Biomed Chromatogr. 2013 Dec;27(12):1775-81. doi: 10.1002/bmc.2993. Epub 2013 Jul 11.

Abstract

A simple and sensitive method based on the combination of solid-phase microextraction (SPME) and high-performance liquid chromatography with ultroviolet detection was developed for the simultaneous determination of clenbuterol, salbutamol and ractopamine in pig samples. Parameters of the SPME procedure affecting extraction efficiency, such as the type of fiber, extraction time, extraction temperature, ion strength, pH of sample and stirring rate, were optimized. The developed method was validated according to the International Conference on Harmonization guidelines. The calibration curves were linear over a range of 0.5-50 µg/L for clenbuterol and ractopamine, and 0.2-20 µg/L for salbutamol. The limits of detection were 0.1 µg/L for clenbuterol, 0.05 µg/L for salbutamol and 0.1 μg/L for ractopamine, respectively. The averages of intra- and inter-day accuracy ranged from 79.8 to 92.4%. The intra-day and inter-day precision were below 9.6% for the three analytes. This method exhibited the advantages of simplicity, rapidity and low solvent consumption, and was suitable for the monitoring of β2 -agonists residue in pig samples.

Keywords: clenbuterol; high-performance liquid chromatography-ultraviolet detection; ractopamine; salbutamol; solid-phase microextraction.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Albuterol / analysis*
  • Albuterol / chemistry
  • Albuterol / isolation & purification
  • Animals
  • Chromatography, High Pressure Liquid / methods*
  • Clenbuterol / analysis*
  • Clenbuterol / chemistry
  • Clenbuterol / isolation & purification
  • Limit of Detection
  • Linear Models
  • Liver / chemistry
  • Muscles / chemistry
  • Phenethylamines / analysis*
  • Phenethylamines / chemistry
  • Phenethylamines / isolation & purification
  • Reproducibility of Results
  • Solid Phase Microextraction / methods*
  • Spectrophotometry, Ultraviolet
  • Swine

Substances

  • Phenethylamines
  • ractopamine
  • Albuterol
  • Clenbuterol