Improved diffuse fluorescence flow cytometer prototype for high sensitivity detection of rare circulating cells in vivo

J Biomed Opt. 2013 Jul;18(7):077002. doi: 10.1117/1.JBO.18.7.077002.

Abstract

Detection and enumeration of rare circulating cells in mice are important problems in many areas of preclinical biomedical research. Recently, we developed a new method termed "diffuse fluorescence flow cytometry" (DFFC) that uses diffuse photons to increase the blood sampling volume and sensitivity versus existing in vivo flow cytometry methods. In this work, we describe a new DFFC prototype with approximately an order-of-magnitude improvement in sensitivity compared to our previous work. This sensitivity improvement is enabled by a number of technical innovations, which include a method for the removal of motion artifacts (allowing interrogation of mouse hindlegs that was less optically attenuating versus the tail) and improved collection optics and signal preamplification. We validated our system first in limb mimicking optical flow phantoms with fluorescent microspheres and then in nude mice with fluorescently labeled mesenchymal stem cells at injected concentrations of 5×103 cells/mL. In combination, these improvements resulted in an overall cell counting sensitivity of about 1 cell/mL or better in vivo.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Flow Cytometry / methods*
  • Fluorescent Dyes / chemistry
  • Mesenchymal Stem Cells
  • Mice
  • Mice, Nude
  • Microspheres
  • Optical Imaging / methods*
  • Phantoms, Imaging
  • Sensitivity and Specificity
  • Signal-To-Noise Ratio

Substances

  • Fluorescent Dyes