Assessment of pathogenic bacteria using periodic actuation

Lab Chip. 2013 Aug 21;13(16):3192-8. doi: 10.1039/c3lc50411e. Epub 2013 Jun 27.

Abstract

A new analytical platform for the assessment of pathogenic bacteria is presented. It is based on a robust technology which is able to amplify the signal to noise ratio providing fast and sensitive detection of target pathogenic bacteria. The system uses a custom made AC electrical impedance analyser to measure, using a lab on a chip platform, the oscillations of magnetically labelled analytes when applying a periodic magnetic field. The concentration of pathogenic Escherichia coli O157:H7 chosen as bacterial model was determined based on the amplitude of the electrical impedance oscillations at a selected AC frequency. The analytical platform provides a limit of detection of 10(2) cells ml(-1), has a fast analysis time, and is amenable for the detection of other target cells. The system has simple design suitable for portability and automated operation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrodes
  • Escherichia coli O157 / isolation & purification*
  • Lab-On-A-Chip Devices*
  • Magnetic Fields
  • Magnets / chemistry
  • Microspheres
  • Time Factors