[Determination of1-bromopropane in the workplace air by GC-FID]

Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2013 Jun;31(6):467-9.
[Article in Chinese]

Abstract

Objective: To establish a method for determination 1-bromopropane (1-BP) in workplace air by gas chromatography-flame ionization detector (GC-FID).

Method: 1-bromopropane in workplace air was collected with activated charcoal tube, then desorbed by carbon disulfide and determined by GC-FID. 1-bromopropane was quantitatively measured using retention time and peak area.

Results: Linear regression formula was Y = 3353.4x-10064 in a range of 2.50 ∼ 500.00 µg/ml with regression coefficient R = 0.9998. Detection limit was 0.25 µg/ml and the lowest detection concentration of 1-brmopropane in air was 0.14 mg/m(3) (at air volume 1.8L). The mean recoveries of 1-BP were between 96.8% and 102.6%, and relative standard deviation of inter and intra-assay was less than 10%. The average desorption efficiencies were between 93.2% and 104.4%. The samples in activated charcoal tube could be stably stored for 5 days at room temperature.

Conclusion: The method could be feasible to determine 1-bromopropane in workplace air.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air / analysis
  • Air Pollutants, Occupational / analysis*
  • Chromatography, Gas / methods*
  • Hydrocarbons, Brominated / analysis
  • Workplace

Substances

  • Air Pollutants, Occupational
  • Hydrocarbons, Brominated
  • 1-bromopropane