[A practical procedure to improve the accuracy of radiochromic film dosimetry: a integration with a correction method of uniformity correction and a red/blue correction method]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2013 Jun;69(6):617-31. doi: 10.6009/jjrt.2013_jsrt_69.6.617.
[Article in Japanese]

Abstract

It has been reported that the light scattering could worsen the accuracy of dose distribution measurement using a radiochromic film. The purpose of this study was to investigate the accuracy of two different films, EDR2 and EBT2, as film dosimetry tools. The effectiveness of a correction method for the non-uniformity caused from EBT2 film and the light scattering was also evaluated. In addition the efficacy of this correction method integrated with the red/blue correction method was assessed. EDR2 and EBT2 films were read using a flatbed charge-coupled device scanner (EPSON 10000G). Dose differences on the axis perpendicular to the scanner lamp movement axis were within 1% with EDR2, but exceeded 3% (Maximum: +8%) with EBT2. The non-uniformity correction method, after a single film exposure, was applied to the readout of the films. A corrected dose distribution data was subsequently created. The correction method showed more than 10%-better pass ratios in dose difference evaluation than when the correction method was not applied. The red/blue correction method resulted in 5%-improvement compared with the standard procedure that employed red color only. The correction method with EBT2 proved to be able to rapidly correct non-uniformity, and has potential for routine clinical IMRT dose verification if the accuracy of EBT2 is required to be similar to that of EDR2. The use of red/blue correction method may improve the accuracy, but we recommend we should use the red/blue correction method carefully and understand the characteristics of EBT2 for red color only and the red/blue correction method.

Publication types

  • English Abstract

MeSH terms

  • Film Dosimetry / instrumentation
  • Film Dosimetry / methods*
  • Quality Assurance, Health Care
  • Radiation Dosage
  • Radiotherapy, Intensity-Modulated / methods