A new versatile underground gamma-ray spectrometry system

Appl Radiat Isot. 2013 Nov:81:81-6. doi: 10.1016/j.apradiso.2013.03.079. Epub 2013 Apr 6.

Abstract

The newest development in IRMM's underground analytical facility is a large lead shield lined with copper that is versatile and can host several detectors of different types. The characteristics and the background performance of the shield are described for four different detector configurations involving HPGe-detectors and NaI-detectors. The shield has been designed to swap detectors, while still maintaining a low background. This enables testing of detectors for other experiments and optimisation of detection limits for specific radionuclides in different projects.

Keywords: Compton suppression; HPGe-detector; Low-level gamma-ray spectrometry; Shielding; Underground laboratory.

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Radiation Dosage
  • Radiation Protection / instrumentation*
  • Radioisotopes / analysis*
  • Radiometry / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Gamma / instrumentation*

Substances

  • Radioisotopes