Frameless high dose rate stereotactic lung radiotherapy: intrafraction tumor position and delivery time

Radiother Oncol. 2013 Jun;107(3):419-22. doi: 10.1016/j.radonc.2013.04.019. Epub 2013 May 23.

Abstract

Intrafraction change in tumor position (Δ) was evaluated for stereotactic lung radiotherapy delivered with flattening filter free volumetric modulated arc therapy. In 140 fractions from 32 patients mean Δ (±SD) was -0.7±1.4 mm (vertical), -0.7±1.3 mm (longitudinal) and +0.2±1.2 mm (lateral) with mean vector 2.1±1.2 mm. Mean delivery time was 4.4±3.4 min (mean beam-on 1.9±0.4 min).

Keywords: Flattening filter free; Intrafraction motion; Lung cancer; SBRT.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cone-Beam Computed Tomography
  • Dose Fractionation, Radiation
  • Female
  • Humans
  • Lung Neoplasms / diagnostic imaging
  • Lung Neoplasms / pathology
  • Lung Neoplasms / surgery*
  • Male
  • Radiosurgery / methods*
  • Time Factors