Long-term left ventricular support in patients with a mechanical aortic valve

Scand Cardiovasc J. 2013 Aug;47(4):236-9. doi: 10.3109/14017431.2013.795655. Epub 2013 May 22.

Abstract

Objectives: The presence of a mechanical prosthesis has been regarded as an increased risk of thromboembolic complications and as a relative contraindication for a left ventricular assist device (LVAD). Five patients in our center had a mechanical aortic valve at the time of device implantation and were studied regarding thromboembolic complications.

Design: Five patients operated upon with an LVAD (1 HeartMate I™, 4 HeartMate II™) between 2002 and 2011 had a mechanical aortic valve at the time of implantation. The first patient had a patch closure of the aortic valve. In four patients, the prosthesis was left in place. Anticoagulants included aspirin, warfarin, and clopidogrel.

Results: The average and accumulated treatment times were 150 and 752 days, respectively. Three of the five patients showed early signs of valve thrombosis on echo with concomitant valve dysfunction. Four patients were transplanted without thromboembolic events during pump treatment. One patient died from a hemorrhagic stroke after 90 days on the LVAD.

Conclusions: The strategy of leaving a mechanical heart valve in place at the time of LVAD implantation in five patients led to valvular thrombosis in three but did not provoke embolic events. It increased the complexity of postoperative anticoagulation.

MeSH terms

  • Adult
  • Anticoagulants / therapeutic use
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Cerebral Hemorrhage / etiology
  • Female
  • Heart Failure / diagnosis
  • Heart Failure / physiopathology
  • Heart Failure / surgery*
  • Heart Transplantation
  • Heart Valve Diseases / diagnosis
  • Heart Valve Diseases / physiopathology
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis*
  • Heart-Assist Devices*
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Thrombosis / etiology
  • Time Factors
  • Treatment Outcome
  • Ventricular Function, Left*

Substances

  • Anticoagulants