Non-invasive, minute-to-minute estimates of systemic arterial pressure and pulse pressure variation using radial artery tonometry

J Med Eng Technol. 2013 Apr;37(3):197-202. doi: 10.3109/03091902.2013.774443.

Abstract

The Tensys T-line uses tonometry to reproduce the arterial blood pressure tracing non-invasively. The purpose of this study was to assess the agreement between estimates of the T-line and an intra-arterial catheter (for both mean arterial pressure [MAP] and pulse pressure variation [PPV]) in the setting of spine surgery. Continuous blood pressure data were collected for 7507 minutes from 25 patients. Five increasingly aggressive T-line filters were applied. The mean bias for mean, diastolic and systolic blood pressure ranged from 3.4-6.4, 3.1-7.1 and 0.1-0.8 mmHg and 6.5-11.8% for PPV. Ninety-five per cent confidence intervals for mean, diastolic and systolic blood pressure ranged from 24-28, 23.1-24.7 and 33.4-35.6 mmHg for 14-21% for PPV. The limits of agreement preclude the use of the T-line for reliable estimation of MAP or PPV in spine surgery.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Pressure Determination / methods
  • Blood Pressure*
  • Humans
  • Manometry
  • Monitoring, Intraoperative / methods
  • Radial Artery / physiology*