High sensitivity trace gas detection by cantilever-enhanced photoacoustic spectroscopy using a mid-infrared continuous-wave optical parametric oscillator

Opt Express. 2013 Apr 22;21(8):10240-50. doi: 10.1364/OE.21.010240.

Abstract

Highly sensitive cantilever-enhanced photoacoustic detection of hydrogen cyanide and methane in the mid-infrared region is demonstrated. A mid-infrared continuous-wave frequency tunable optical parametric oscillator was used as a light source in the experimental setup. Noise equivalent detection limits of 190 ppt (1 s) and 65 ppt (30 s) were achieved for HCN and CH(4), respectively. The normalized noise equivalent absorption coefficient is 1.8 × 10(-9) W cm(-1) Hz(-1/2).

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Gases / analysis*
  • Gases / chemistry
  • Micro-Electrical-Mechanical Systems / instrumentation*
  • Microchemistry / instrumentation*
  • Oscillometry / instrumentation*
  • Photoacoustic Techniques / instrumentation*
  • Spectrophotometry, Infrared / instrumentation*

Substances

  • Gases