Overdetermined broadband spectroscopic Mueller matrix polarimeter designed by genetic algorithms

Opt Express. 2013 Apr 8;21(7):8753-62. doi: 10.1364/OE.21.008753.

Abstract

This paper reports on the design and implementation of a liquid crystal variable retarder based overdetermined spectroscopic Mueller matrix polarimeter, with parallel processing of all wavelengths. The system was designed using a modified version of a recently developed genetic algorithm [Letnes et al. Opt. Express 18, 22, 23095 (2010)]. A generalization of the eigenvalue calibration method is reported that allows the calibration of such overdetermined polarimetric systems. Out of several possible designs, one of the designs was experimentally implemented and calibrated. It is reported that the instrument demonstrated good performance, with a measurement accuracy in the range of 0.1% for the measurement of air.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Equipment Design
  • Equipment Failure Analysis
  • Liquid Crystals / chemistry*
  • Liquid Crystals / radiation effects
  • Models, Genetic
  • Refractometry / instrumentation*
  • Spectrum Analysis / instrumentation*