Determination of glutathione and cysteine in yeasts by hydrophilic interaction liquid chromatography followed by on-line postcolumn derivatization

J Sep Sci. 2013 Jun;36(12):1877-82. doi: 10.1002/jssc.201300202. Epub 2013 May 18.

Abstract

In the present study, we report a new method for the determination of two primary thiols, cysteine (CYS) and glutathione (GSH), by hydrophilic interaction LC. The polar analytes are separated isocratically using a mobile phase consisting of 65% acetonitrile/35% ammonium acetate (15 mmol/L, pH 2.0) and are detected at 285 nm following on-line postcolumn derivatization by the thiol-selective reagent methyl propiolate. The main figures of merit included linearity in the range of 5-200 μmol/L and an LOD 0.6 μmol/L for both compounds. The absence of matrix effect allowed the determination of CYS and GSH in various yeast samples. GSH was present in most of the samples at levels ranging between 0.9 and 3.1 mg/g, whereas CYS was determined in only one sample at a significantly lower concentration. In terms of accuracy, the percent recoveries ranged between 91.2 and 105.6% for GSH, and 91.6 and 106.9% for CYS.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Automation
  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Cysteine / analysis*
  • Glutathione / analysis*
  • Hydrophobic and Hydrophilic Interactions
  • Yeasts / chemistry*

Substances

  • Glutathione
  • Cysteine