Demonstration of a plenoptic microscope based on laser optical feedback imaging

Opt Express. 2013 Mar 25;21(6):7294-303. doi: 10.1364/OE.21.007294.

Abstract

A new kind of plenoptic imaging system based on Laser Optical Feedback Imaging (LOFI) is presented and is compared to another previously existing device based on microlens array. Improved photometric performances, resolution and depth of field are obtained at the price of a slow point by point scanning. Main properties of plenoptic microscopes such as numerical refocusing on any curved surface or aberrations compensation are both theoretically and experimentally demonstrated with a LOFI-based device.

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Feasibility Studies
  • Feedback
  • Lasers*
  • Microscopy / instrumentation*
  • Photometry / instrumentation*