The Monorail Piccolino coronary angioplasty balloon catheter (MBC) was evaluated on 118 patients at two centers. Technical success was achieved in 110 patients (93%). Time for catheter exchange and total fluoroscopy time were significantly lower for the Monorail catheter than with standard equipment (exchange time 97 vs. 170 seconds P less than .05 and fluoroscopy time 17 vs. 88 seconds P less than .001). The advantages of rapid exchange and the ability of utilize 2 Monorail balloon catheters through one 9F guiding catheter for simultaneous inflations allowed for maximal flexibility in treating patients with bifurcation lesions. The double wire approach utilizing one Monorail balloon catheter with a 7F guiding catheter was also technically successful. The Monorail Piccolino balloon catheter has unique features that allow for greater ease of operator use, rapid catheter exchange, and optimal angiographic visualization. It is felt that this catheter design provides distinct advantages over standard angioplasty equipment.