A rapid liquid chromatography tandem mass spectrometry-based method for measuring propranolol on dried blood spots

J Pharm Biomed Anal. 2013 May 5:78-79:34-8. doi: 10.1016/j.jpba.2013.01.034. Epub 2013 Feb 4.

Abstract

Propranolol, a non-selective beta blocker drug, is used in young infants and newborns for treating several heart diseases; its pharmacokinetics has been extensively evaluated in adult patients using extrapolation to treat pediatric population. The purpose of the present study was to develop and validate a method to measure propranolol levels in dried blood spots. The analysis was performed by using liquid chromatography/tandem mass spectrometry operating in multiple reaction monitoring mode. The calibration curve in matrix was linear in the concentration range of 2.5-200 μg/L with correlation coefficient r=0.9996. Intra-day and inter-day precisions and biases were less than 8.0% (n=10) and 11.5% (n=10) respectively. The recoveries ranged from 94 to 100% and the matrix effect did not result in a severe signal suppression. Propranolol on dried blood spot showed a good stability at three different temperatures for one month. This paper describes a micromethod for measuring propranolol levels on dried blood spot, which determines a great advantage in neonates or young infants during pharmacokinetic studies because of less invasive sampling and small blood volume required.

Publication types

  • Validation Study

MeSH terms

  • Adrenergic beta-Antagonists / blood*
  • Calibration
  • Chromatography, Liquid / methods*
  • Humans
  • Propranolol / blood*
  • Reproducibility of Results
  • Tandem Mass Spectrometry / methods*

Substances

  • Adrenergic beta-Antagonists
  • Propranolol