Outbreak of multidrug-resistant Pseudomonas aeruginosa infection following urodynamic studies traced to contaminated transducer

J Hosp Infect. 2013 Apr;83(4):344-6. doi: 10.1016/j.jhin.2013.01.003. Epub 2013 Feb 16.

Abstract

We report a nosocomial outbreak of urinary tract infection by extremely drug resistant Pseudomonas aeruginosa, susceptible only to colistin. Infection in three patients followed urodynamic studies. Two of the three patients were children, one of whom also developed urosepsis. The investigation led to detection of contaminated pressure transducers. Genotyping confirmed that patient and transducer isolates were identical. These transducers were not labelled as 'single use only' despite the possibility that contaminated urine may reflux and mix with the fluid in the device. The issue of re-usable versus single-use urodynamic devices is discussed.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Child
  • Cross Infection / epidemiology*
  • Cross Infection / microbiology
  • Disease Outbreaks*
  • Drug Resistance, Multiple, Bacterial*
  • Equipment and Supplies / microbiology*
  • Genotype
  • Humans
  • Male
  • Molecular Typing
  • Pseudomonas Infections / epidemiology*
  • Pseudomonas Infections / microbiology
  • Pseudomonas aeruginosa / classification
  • Pseudomonas aeruginosa / drug effects*
  • Pseudomonas aeruginosa / genetics
  • Pseudomonas aeruginosa / isolation & purification
  • Urinary Tract Infections / epidemiology*
  • Urinary Tract Infections / microbiology