Non-invasive embolus trap using histotripsy-an acoustic parameter study

Ultrasound Med Biol. 2013 Apr;39(4):611-9. doi: 10.1016/j.ultrasmedbio.2012.11.026. Epub 2013 Feb 13.

Abstract

Free-flowing particles in a blood vessel were observed to be attracted, trapped and eroded by a histotripsy bubble cloud. This phenomenon may be used to develop a non-invasive embolus trap (NET) to prevent embolization. This study investigates the effect of acoustic parameters on the trapping ability of the NET generated by a focused 1.063 MHz transducer. The maximum trapping velocity, defined by the maximum mean fluid velocity at which a 3-4 mm particle trapped in a 6 mm diameter vessel phantom, increased linearly with peak negative pressure (P-) and increased as the square root of pulse length and pulse repetition frequency (PRF). At 19.9 MPa P-, 1000 Hz PRF and 10 cycle pulse length, a 3 mm clot-mimicking particle could remain trapped under a background velocity of 9.7 cm/s. Clot fragments treated by NET resulted in debris particles <75 μm. These results will guide the appropriate selection of NET parameters.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Biomimetics / methods
  • Embolism / physiopathology*
  • Embolism / prevention & control*
  • Humans
  • Lithotripsy / methods*
  • Mechanical Thrombolysis / methods*