Multimodal visibility (radiography, computed tomography, and magnetic resonance imaging) of microspheres for transarterial embolization tested in porcine kidneys

Invest Radiol. 2013 Apr;48(4):213-22. doi: 10.1097/RLI.0b013e31827f6598.

Abstract

Objective: The objective of this study was to test multimodal visibility (radiography, computed tomography [CT], and magnetic resonance imaging [MRI]) of microspheres for transarterial embolization in porcine kidneys.

Materials and methods: Currently available embolization particles (microspheres) were modified. A dense x-ray material (barium sulfate) was added to create visibility for radiography and CT. A magnetic substance (iron oxide) was additionally added to create visibility for MRI. This chemical modification was performed for particles with sizes of 100 ± 25 and 700 ± 50 μm. Three different prototypes per size class (samples A, B, and C) resulted, each with a different degree of barium sulfate but with the same degree of iron oxide. The currently available embolization particles with sizes of 100 ± 25 and 700 ± 50 μm were used as controls (sample control). Eight renal arteries of 4 pigs were embolized. Study end points were size distribution evaluated in vitro as well as qualitative and quantitative particle visibility evaluated in vivo.

Results: The size distribution of the particles with a size of 100 ± 25 μm was between 96 ± 11 μm for sample A and 102 ± 13 μm for the sample control without significant differences (n.s.). The size distribution of the particles with a size of 700 ± 50 μm was between 691 ± 20 μm for sample A and 716 ± 34 μm for sample C without significant differences (n.s.). For radiography, the particles with sizes of 100 ± 25 and 700 ± 50 μm for samples A, B, and C were definitely visible during the embolization. The sample control was definitely not visible. For CT and MRI (T1-weighted [T1w] and T2-weighted [T2w]), the particles with sizes of 100 ± 25 and 700 ± 50 μm for samples A, B, and C were definitely visible after the embolization. The sample control was definitely not visible. For CT, the signal-to-noise ratio for samples A, B, and C increased significantly after the embolization (eg, sample A, particles with a size of 100 ± 25 μm: 66.5% ± 23.7%, P < 0.05). The signal-to-noise ratio for the sample control did not change after the embolization (eg, sample control, particles with a size of 700 ± 25 μm: -0.2% ± 15.2%, n.s.). For MRI (T1w and T2w), the signal-to-noise ratio for samples A, B, and C decreased significantly after the embolization (eg, sample B, particles with a size of 700 ± 50 μm, T1w: -72.9% ± 6.6%; P < 0.05). The signal-to-noise ratio for the sample control did not change after the embolization (eg, sample control, particles with a size of 100 ± 25 μm, T2w: 6.2% ± 16.1%, n.s.).

Conclusions: In this study, the chemical modification of the currently available microspheres for transarterial embolization resulted in a size distribution comparable with the currently available microspheres and created multimodal visibility for radiography, CT, and MRI.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Barium Sulfate / administration & dosage
  • Contrast Media / administration & dosage
  • Dextrans / administration & dosage
  • Embolization, Therapeutic / methods*
  • Female
  • Kidney*
  • Magnetic Resonance Imaging / methods*
  • Magnetite Nanoparticles / administration & dosage
  • Microspheres*
  • Swine
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Dextrans
  • Magnetite Nanoparticles
  • Barium Sulfate
  • ferumoxides