Delayed first otoacoustic emissions test decreases failure on neonatal hearing screening after caesarean delivery

Acta Paediatr. 2013 May;102(5):e194-9. doi: 10.1111/apa.12175. Epub 2013 Feb 12.

Abstract

Aim: Caesarean delivery (CD) was associated with a 3.2-fold higher failure on 1st otoacoustic emissions (OAE) hearing test. We aimed to verify whether postponing 1st OAE beyond 48 h in CD infants decreases hearing screening failure.

Methods: We compared two groups of CD infants as to failure on 1st OAE test: early-1st OAE (n = 560): 1st OAE at 12- to 48-h-olds and late-1st OAE (n = 566): 1st OAE at 48- to 132-h-olds.

Results: Compared with early-1st OAE group, the failure rate among late-1st OAE infants was significantly sixfold lower (20.5% vs. 3.4%), with sixfold lower need for repeated tests: 205 vs. 34 tests/1000 CD neonates (p < 0.001). The failure rate decreased with increasing age in both groups (p < 0.001). Univariate analysis: timing of 1st OAE (late vs. early) was significantly associated with failure on 1st OAE. Multivariable analysis: late (48-132 h) 1st OAE test was associated with a 7.7-fold lower risk for failure of 1st OAE, OR (95% CI): 0.13 (0.08-0.21).

Conclusion: Among CD infants, the risk for failure in late-1st OAE group (>48 h) was 7.7-fold lower, with a sixfold lower need for repeated hearing tests. Delaying 1st OAE in CD infants beyond 48 h of age (preferably between 48 and 132 h) decreases neonatal OAE screening failure.

Publication types

  • Comparative Study

MeSH terms

  • Cesarean Section / adverse effects*
  • False Positive Reactions
  • Female
  • Hearing Tests
  • Humans
  • Infant, Newborn
  • Male
  • Neonatal Screening*
  • Otoacoustic Emissions, Spontaneous*