Effect of anterior-posterior and internal-external motion restraint during knee wear simulation on a posterior stabilised knee design

J Biomech. 2013 Feb 1;46(3):491-7. doi: 10.1016/j.jbiomech.2012.10.017. Epub 2012 Nov 13.

Abstract

The objective of our study was to examine the effect of biphaseal AP translation and IE rotation restraint, using a system defined specifically for posterior stabilised knee designs, on wear, kinematics and particle release in comparison to linear motion restraint as required by the established ISO 14243-1:2002(E) protocol. In the ISOlinear groups, an AP motion restraint of 30 N/mm and an IE rotation restraint of 0.6 Nm/° were applied in the knee wear simulation. In the ISOgap biphaseal groups with PCL sacrificing implants, the restraining AP force was zero in a ±2.5mm range with, externally, a constant of 9.3N/mm applied proportionally to the AP translation of the tibia plateau, whereas the restraining IE torque was zero in a ±6° range with, externally, a constant of 0.13 Nm/° applied proportionally to the IE rotation of the tibia plateau. Using the ISOgap biphaseal protocol on a posterior stabilised knee design, we found an increase of 41% in AP translation and of 131% in IE rotation, resulting in a 3.2-fold higher wear rate compared to the results obtained using the ISOlinear protocol. Changes in AP translation and IE rotation ligament motion restraints have a high impact on knee joint kinematics and wear behaviour of a fixed bearing posterior stabilised knee design.

MeSH terms

  • Biomechanical Phenomena
  • Humans
  • Knee Prosthesis*
  • Materials Testing / methods*
  • Motion*
  • Prosthesis Design*