Significant change or loss of intraoperative monitoring data: a 25-year experience in 12,375 spinal surgeries

Spine (Phila Pa 1976). 2013 Jan 15;38(2):E101-8. doi: 10.1097/BRS.0b013e31827aafb9.

Abstract

Study design: Retrospective.

Objective: The purpose of this study was to report the spectrum of intraoperative events responsible for a loss or significant change in intraoperative monitoring (IOM) data.

Summary of background data: The efficacy of spinal cord/nerve root monitoring is demonstrated in a large, single institution series of patients, involving all levels of the spinal column (occiput to sacrum) and all spinal surgical procedures.

Methods: Multimodality IOM included somatosensory-evoked potentials, descending neurogenic-evoked potentials, neurogenic motor-evoked potentials, and spontaneous and triggered electromyography. A total of 12,375 patients who underwent surgery for spinal pathology between January 1985 and December 2010 were reviewed. There were 59.3% female patients (7178) and 40.7% male patients (5197). Procedures by spinal level were as follows: cervical 29.7% (3671), thoracic/thoracolumbar 45.4% (5624), and lumbosacral 24.9% (3080). Age at the time of surgery was as follows: older than 18 years, 72.7% (242/8993) and younger than 18 years, 27.3% (144/3382). A total of 77.8% (9633) patients underwent primary surgical procedures and 22.2% (2742) patients underwent revision surgical procedures.

Results: A total of 406 instances of IOM data change/loss occurred in 386 of 12,375 (3.1%) patients. Causes for data degradation/loss included the following: instrumentation (n = 131), positioning (n = 85), correction (n = 56), systemic (n = 49), unknown (n = 24), and focal spinal cord compression (n = 15). Data loss/change was seen in revision (6.1%/167 patients) surgical procedures more commonly than in primary procedures (2.3%/219 patients; P < 0.0001). Data improvement was demonstrated by 88.7% (n = 360) after intervention versus 11.3% (n = 46) with no improvement in IOM data. One patient with improved data after intervention versus 14 with no improvement despite intervention had a permanent neurological deficit (P < 0.0001).

Conclusion: IOM data identified 386 (3.1%) patients with loss/degradation of data in 12,375 spinal surgical procedures. Fortunately, in 93.3% of patients, intervention led to data recovery and no neurological deficits. Reduction from a potential (worst-case scenario) 3.1% (386) of patients with significant change/loss of IOM data to a permanent neurological deficit rate of 0.12% (15) patients was achieved (P < 0.0001), thus confirming efficacy of IOM.

MeSH terms

  • Adolescent
  • Adult
  • Child
  • Electromyography
  • Evoked Potentials, Motor / physiology
  • Evoked Potentials, Somatosensory / physiology
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Information Storage and Retrieval
  • Intraoperative Complications
  • Intraoperative Period
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / methods*
  • Neurosurgical Procedures
  • Practice Patterns, Physicians' / statistics & numerical data*
  • Retrospective Studies
  • Spinal Cord / physiology
  • Spinal Cord / surgery*
  • Spinal Cord Injuries / prevention & control
  • Spinal Diseases / physiopathology
  • Spinal Diseases / surgery*