Effect of inhomogeneity in a patient's body on the accuracy of the pencil beam algorithm in comparison to Monte Carlo

Phys Med Biol. 2012 Nov 21;57(22):7673-88. doi: 10.1088/0031-9155/57/22/7673. Epub 2012 Nov 2.

Abstract

The pencil beam algorithm (PBA) is reasonably accurate and fast. It is, therefore, the primary method used in routine clinical treatment planning for proton radiotherapy; still, it needs to be validated for use in highly inhomogeneous regions. In our investigation of the effect of patient inhomogeneity, PBA was compared with Monte Carlo (MC). A software framework was developed for the MC simulation of radiotherapy based on Geant4. Anatomical sites selected for the comparison were the head/neck, liver, lung and pelvis region. The dose distributions calculated by the two methods in selected examples were compared, as well as a dose volume histogram (DVH) derived from the dose distributions. The comparison of the off-center ratio (OCR) at the iso-center showed good agreement between the PBA and MC, while discrepancies were seen around the distal fall-off regions. While MC showed a fine structure on the OCR in the distal fall-off region, the PBA showed smoother distribution. The fine structures in MC calculation appeared downstream of very low-density regions. Comparison of DVHs showed that most of the target volumes were similarly covered, while some OARs located around the distal region received a higher dose when calculated by MC than the PBA.

Publication types

  • Comparative Study

MeSH terms

  • Algorithms*
  • Humans
  • Monte Carlo Method*
  • Organs at Risk / radiation effects
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*