A novel instrumentation circuit for electrochemical measurements

Sensors (Basel). 2012;12(7):9687-96. doi: 10.3390/s120709687. Epub 2012 Jul 17.

Abstract

In this paper, a novel signal processing circuit which can be used for the measurement of H(+) ion and urea concentration is presented. A potentiometric method is used to detect the concentrations of H(+) ions and urea by using H(+) ion-selective electrodes and urea electrodes, respectively. The experimental data shows that this measuring structure has a linear pH response for the concentration range within pH 2 and 12, and the dynamic range for urea concentration measurement is in the range of 0.25 to 64 mg/dL. The designed instrumentation circuit possesses a calibration function and it can be applied to different sensing electrodes for electrochemical analysis. It possesses the advantageous properties of being multi-purpose, easy calibration and low cost.

Keywords: biosensor; electrochemical measurement; instrumentation; multi-purpose.

Publication types

  • Research Support, Non-U.S. Gov't