Incidence and clinical impact of stent fracture after everolimus-eluting stent implantation

Circ Cardiovasc Interv. 2012 Oct;5(5):663-71. doi: 10.1161/CIRCINTERVENTIONS.112.969238. Epub 2012 Sep 25.

Abstract

Background: Stent fracture (SF) after drug-eluting stent implantation has recently become an important concern because of its potential association with in-stent restenosis and stent thrombosis. However, the incidence and clinical impact of SF after everolimus-eluting stent implantation remain unclear.

Methods and results: A total of 1035 patients with 1339 lesions undergoing everolimus-eluting stent implantation and follow-up angiography 6 to 9 months after index procedure were analyzed. SF was defined as complete or partial separation of the stent, as assessed by plain fluoroscopy or intravascular ultrasound during follow-up. We assessed the rates of SF and major adverse cardiac events, defined as cardiac death, myocardial infarction, stent thrombosis, and clinically driven target lesion revascularization within 9 months. SF was observed in 39 of 1339 lesions (2.9%) and in 39 of 1035 patients (3.8%). Ostial stent location and lesions with hinge motion, tortuosity, or calcification were independent predictors of SF. The rate of myocardial infarction and target lesion revascularization were significantly higher in the SF group than in the non-SF group (5.1% versus 0.4%; P=0.018 and 25.6% versus 2.0%; P<0.001, respectively). Stent thrombosis was more frequently observed in the SF group than in the non-SF group (5.1% versus 0.4%; P=0.018). Major adverse cardiac events within 9 months were significantly higher in the SF group than in the non-SF group (25.6% versus 2.3%; P<0.001).

Conclusions: SF after everolimus-eluting stent implantation occurs in 2.9% of lesions and is associated with higher rate of major adverse cardiac events, driven by higher target lesion revascularization and stent thrombosis.

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Angioplasty, Balloon, Coronary / mortality
  • Cardiovascular Agents / administration & dosage*
  • Chi-Square Distribution
  • Coronary Angiography
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Thrombosis / diagnosis
  • Coronary Thrombosis / etiology
  • Coronary Thrombosis / mortality
  • Drug-Eluting Stents*
  • Everolimus
  • Female
  • Humans
  • Incidence
  • Japan
  • Logistic Models
  • Male
  • Middle Aged
  • Multivariate Analysis
  • Myocardial Infarction / diagnosis
  • Myocardial Infarction / etiology
  • Myocardial Infarction / mortality
  • Odds Ratio
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Failure*
  • Risk Factors
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Time Factors
  • Ultrasonography, Interventional
  • Vascular Calcification / diagnosis
  • Vascular Calcification / mortality
  • Vascular Calcification / therapy

Substances

  • Cardiovascular Agents
  • Everolimus
  • Sirolimus