Using left-ventricular-only pacing to eliminate T-wave oversensing in a biventricular implantable cardiac defibrillator

Can J Cardiol. 2013 Feb;29(2):254.e5-7. doi: 10.1016/j.cjca.2012.06.005. Epub 2012 Aug 15.

Abstract

A man aged 75 years and with nonischemic cardiomyopathy had implantation of a biventricular implantable cardiac defibrillator (ICD). Consistent biventricular pacing was limited by intermittent T-wave oversensing (TWOS). A strategy of left-ventricular-only pacing was used to eliminate TWOS. This strategy obviates the need to reduce ventricular sensitivity and thus may be an effective alternative to biventricular pacing complicated by TWOS.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy / methods*
  • Defibrillators, Implantable*
  • Electrocardiography*
  • Equipment Failure
  • Heart Failure / complications
  • Heart Failure / physiopathology
  • Heart Failure / therapy*
  • Humans
  • Male
  • Tachycardia, Ventricular / complications
  • Tachycardia, Ventricular / therapy*