The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study

Heart Rhythm. 2012 Nov;9(11):1789-95. doi: 10.1016/j.hrthm.2012.07.016. Epub 2012 Jul 20.

Abstract

Background: The clinical efficacy of catheter ablation of paroxysmal atrial fibrillation (AF) remains limited by difficulty in achieving durable pulmonary vein isolation (PVI). Suboptimal catheter tip-to-tissue contact force (CF) during lesion delivery is believed to reduce clinical efficacy.

Objective: To determine the relationship between catheter CF during irrigated catheter ablation for AF and clinical recurrences during follow-up.

Methods: Thirty-two patients with paroxysmal AF underwent PVI by using a radiofrequency ablation catheter with a CF sensor integrated at its tip, and they were followed for 12 months. The relationship between the CF and clinical outcomes was determined.

Results: Acute PVI was achieved in 100% of the veins. Thirty-five percent (351 of 1017) of the applications were placed with an average CF of <10 g (low CF). All patients treated with an average CF of <10 g (5 of 5 patients) experienced recurrences, whereas 80% of the patients treated with an average CF of >20 g (8 of 10 patients) were free from AF recurrence at 12 months. The analysis of the average force-time integral showed that 75% of the patients treated with <500 gs were recurrent whereas only 31% of the patients treated with >1000 gs had recurrences at 12 months.

Conclusions: The CF during catheter ablation for AF correlates with clinical outcome. Arrhythmia control is best achieved when ablation lesions are placed with an average CF of >20 g, and clinical failure is universally noted with an average CF of <10 g.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Atrial Fibrillation / surgery*
  • Catheter Ablation / instrumentation*
  • Catheter Ablation / methods
  • Electrocardiography, Ambulatory
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Physiologic / instrumentation*
  • Recurrence
  • Treatment Outcome