A force-sensing microsurgical instrument that detects forces below human tactile sensation

Retina. 2013 Jan;33(1):200-6. doi: 10.1097/IAE.0b013e3182625d2b.

Abstract

Purpose: To test the sensitivity and reproducibility of a 25-gauge force-sensing micropick during microsurgical maneuvers that are below tactile sensation.

Methods: Forces were measured during membrane peeling in a "raw egg" and the chick chorioallantoic membrane models (N = 12) of epiretinal membranes. Forces were also measured during posterior hyaloid detachment and creation of retinal tears during vitrectomy in live rabbits (n = 6).

Results: With the raw egg model, 0.5 ± 0.4 mN of force was detected during membrane peeling. In the chorioallantoic membrane model, delaminating the upper membrane produced 2.8 ± 0.2 mN of force. While intentionally rupturing the lower membrane to simulate a retinal tear, 7.3 ± 0.5 mN (range, 5.1-9.2 mN; P < 0.001) of force was generated while peeling the upper membrane. During vitrectomy, the minimum force that detached the posterior hyaloid was 6.7 ± 1.1 mN, which was similar to the force of 6.4 ± 1.4 mN that caused a retinal tear. The rate of force generation, as indicated by the first derivative of force generation, was 3.4 ± 1.2 mN/second during posterior hyaloid detachment, compared with 7.7 ± 2.4 mN/second during the creation of a retinal tear (P = 0.04).

Conclusion: Force-sensing microsurgical instruments can detect forces below tactile sensation, and importantly, they can distinguish the forces generated during normal maneuvers from those that cause a surgical complication.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Chick Embryo
  • Chorioallantoic Membrane / surgery
  • Disease Models, Animal
  • Epiretinal Membrane / surgery
  • Equipment Design
  • Microsurgery / instrumentation*
  • Ophthalmologic Surgical Procedures / instrumentation*
  • Pressure*
  • Rabbits
  • Reproducibility of Results
  • Retinal Perforations
  • Robotics / instrumentation*
  • Sensation
  • Sensitivity and Specificity
  • Vitrectomy
  • Vitreous Detachment / surgery