Evaluation of aortic valve stenosis using cardiovascular magnetic resonance: comparison of an original semiautomated analysis of phase-contrast cardiovascular magnetic resonance with Doppler echocardiography

Circ Cardiovasc Imaging. 2012 Sep 1;5(5):604-12. doi: 10.1161/CIRCIMAGING.111.971218. Epub 2012 Jul 13.

Abstract

Background: Accurate quantification of aortic valve stenosis (AVS) is needed for relevant management decisions. However, transthoracic Doppler echocardiography (TTE) remains inconclusive in a significant number of patients. Previous studies demonstrated the usefulness of phase-contrast cardiovascular magnetic resonance (PC-CMR) in noninvasive AVS evaluation. We hypothesized that semiautomated analysis of aortic hemodynamics from PC-CMR might provide reproducible and accurate evaluation of aortic valve area (AVA), aortic velocities, and gradients in agreement with TTE.

Methods and results: We studied 53 AVS patients (AVA(TTE)=0.87±0.44 cm(2)) and 21 controls (AVA(TTE)=2.96±0.59 cm(2)) who had TTE and PC-CMR of aortic valve and left ventricular outflow tract on the same day. PC-CMR data analysis included left ventricular outflow tract and aortic valve segmentation, and extraction of velocities, gradients, and flow rates. Three AVA measures were performed: AVA(CMR1) based on Hakki formula, AVA(CMR2) based on continuity equation, AVA(CMR3) simplified continuity equation=left ventricular outflow tract peak flow rate/aortic peak velocity. Our analysis was reproducible, as reflected by low interoperator variability (<4.56±4.40%). Comparison of PC-CMR and TTE aortic peak velocities and mean gradients resulted in good agreement (r=0.92 with mean bias=-29±62 cm/s and r=0.86 with mean bias=-12±15 mm Hg, respectively). Although good agreement was found between TTE and continuity equation-based CMR-AVA (r>0.94 and mean bias=-0.01±0.38 cm(2) for AVA(CMR2), -0.09±0.28 cm(2) for AVA(CMR3)), AVA(CMR1) values were lower than AVA(TTE) especially for higher AVA (mean bias=-0.45±0.52 cm(2)). Besides, ability of PC-CMR to detect severe AVS, defined by TTE, provided the best results for continuity equation-based methods (accuracy >94%).

Conclusions: Our PC-CMR semiautomated AVS evaluation provided reproducible measurements that accurately detected severe AVS and were in good agreement with TTE.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging*
  • Aortic Valve / physiopathology*
  • Aortic Valve Stenosis / diagnosis*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / physiopathology
  • Automation, Laboratory
  • Blood Flow Velocity
  • Case-Control Studies
  • Echocardiography, Doppler*
  • Female
  • Hemodynamics*
  • Humans
  • Magnetic Resonance Imaging, Cine*
  • Male
  • Middle Aged
  • Observer Variation
  • Paris
  • Predictive Value of Tests
  • Reproducibility of Results
  • Severity of Illness Index
  • Ventricular Function, Left
  • Young Adult