A bio-inspired glucose controller based on pancreatic β-cell physiology

J Diabetes Sci Technol. 2012 May 1;6(3):606-16. doi: 10.1177/193229681200600316.

Abstract

Introduction: Control algorithms for closed-loop insulin delivery in type 1 diabetes have been mainly based on control engineering or artificial intelligence techniques. These, however, are not based on the physiology of the pancreas but seek to implement engineering solutions to biology. Developments in mathematical models of the β-cell physiology of the pancreas have described the glucose-induced insulin release from pancreatic β cells at a molecular level. This has facilitated development of a new class of bio-inspired glucose control algorithms that replicate the functionality of the biological pancreas. However, technologies for sensing glucose levels and delivering insulin use the subcutaneous route, which is nonphysiological and introduces some challenges. In this article, a novel glucose controller is presented as part of a bio-inspired artificial pancreas.

Methods: A mathematical model of β-cell physiology was used as the core of the proposed controller. In order to deal with delays and lack of accuracy introduced by the subcutaneous route, insulin feedback and a gain scheduling strategy were employed. A United States Food and Drug Administration-accepted type 1 diabetes mellitus virtual population was used to validate the presented controller.

Results: Premeal and postmeal mean ± standard deviation blood glucose levels for the adult and adolescent populations were well within the target range set for the controller [(70, 180) mg/dl], with a percent time in range of 92.8 ± 7.3% for the adults and 83.5 ± 14% for the adolescents.

Conclusions: This article shows for the first time very good glucose control in a virtual population with type 1 diabetes mellitus using a controller based on a subcellular β-cell model.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adolescent
  • Adult
  • Algorithms
  • Biosensing Techniques / instrumentation
  • Blood Chemical Analysis / instrumentation
  • Blood Glucose / drug effects*
  • Blood Glucose / metabolism
  • Computer Simulation
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / diagnosis
  • Diabetes Mellitus, Type 1 / drug therapy*
  • Diabetes Mellitus, Type 1 / physiopathology
  • Drug Administration Schedule
  • Equipment Design
  • Humans
  • Hypoglycemia / blood
  • Hypoglycemia / chemically induced
  • Hypoglycemia / prevention & control
  • Hypoglycemic Agents / administration & dosage*
  • Hypoglycemic Agents / adverse effects
  • Infusions, Parenteral
  • Insulin / administration & dosage*
  • Insulin / adverse effects
  • Insulin Infusion Systems*
  • Insulin-Secreting Cells / drug effects*
  • Insulin-Secreting Cells / metabolism
  • Models, Biological*
  • Monitoring, Ambulatory / instrumentation
  • Pancreas, Artificial*
  • Predictive Value of Tests
  • Reproducibility of Results
  • Treatment Outcome
  • Young Adult

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin