Corrosion resistance, chemistry, and mechanical aspects of Nitinol surfaces formed in hydrogen peroxide solutions

J Biomed Mater Res B Appl Biomater. 2012 Aug;100(6):1490-9. doi: 10.1002/jbm.b.32717. Epub 2012 Jun 12.

Abstract

Ti oxides formed naturally on Nitinol surfaces are only a few nanometers thick. To increase their thickness, heat treatments are explored. The resulting surfaces exhibit poor resistance to pitting corrosion. As an alternative approach to accelerate surface oxidation and grow thicker oxides, the exposure of Nitinol to strong oxidizing H(2)O(2) aqueous solutions (3 and 30%) for various periods of time was used. Using X-Ray Photoelectron Spectroscopy (XPS) and Auger spectroscopy, it was found that the surface layers with variable Ti (6-15 at %) and Ni (5-13 at %) contents and the thickness up to 100 nm without Ni-enriched interfaces could be formed. The response of the surface oxides to stress in superelastic regime of deformations depended on oxide thickness. In the corrosion studies performed in both strained and strain-free states using potentiodynamic and potentiostatic polarizations, the surfaces treated in H(2)O(2) showed no pitting in corrosive solution that was assigned to higher chemical homogeneity of the surfaces free of secondary phases and inclusions that assist better biocompatibility of Nitinol medical devices.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Alloys / chemistry*
  • Corrosion
  • Hydrogen Peroxide / chemistry*
  • Materials Testing*
  • Photoelectron Spectroscopy
  • Titanium / chemistry*

Substances

  • Alloys
  • titanium dioxide
  • nitinol
  • Hydrogen Peroxide
  • Titanium