Accuracy of intraocular lens power calculations in eyes with axial length <22.00 mm

Clin Exp Ophthalmol. 2012 Dec;40(9):855-62. doi: 10.1111/j.1442-9071.2012.02810.x. Epub 2012 Jul 11.

Abstract

Background: To assess the accuracy of Haigis, Holladay 1, Hoffer Q and SRK/T formulae in eyes with axial length of <22.00 mm.

Design: Retrospective comparative analysis.

Participants: 163 eyes of 97 patients undergoing phacoemulsification and intraocular lens (IOL) implantation.

Methods: Ocular biometry was performed using IOLMaster laser interferometry. Predicted refractive outcomes before and after lens constant adjustment were compared to actual refractive outcomes.

Main outcome measures: Mean prediction (ME) and mean absolute errors (MAE) with standard deviations (±SD).

Results: Mean preoperative spherical equivalent was +5.44D ± 1.97D. Mean axial length was 21.20 mm ± 0.60 mm. Using standard IOL constants the MAE for Hoffer Q (0.62D, ±0.52D) and Holladay 1 (0.66D ± 0.52D) were significantly lower than SRK/T (MAE 0.91D ± 0.64D; P = <0.0005 and P = 0.001 respectively), but not Haigis (MAE 0.82D ± 0.83D, P = 0.071 and 0.22 respectively). MAEs for all formulae were significantly reduced by IOL constant adjustment (all P = <0.001). Following this there was no statistically significant difference in MAEs between formulae (range 0.50-0.57D, P = 0.57). Increasing MAE was significantly associated with reducing axial length and increasing IOL power for all formulae. For bilateral cases, prediction errors between eyes were significantly correlated across all formulae (all P = <0.0001) and explained 32-42% of the variance in prediction error between eyes.

Conclusions: Prediction of postoperative refraction in patients with short axial lengths is challenging and at the limit of current, popular IOL formulae. There is now a clear need for prospective studies to assess latest generation IOL formulae such as Holladay 2 or Olsen in small eyes.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Axial Length, Eye / abnormalities*
  • Biometry
  • Female
  • Humans
  • Interferometry
  • Lasers
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Optics and Photonics*
  • Phacoemulsification
  • Refraction, Ocular / physiology
  • Reproducibility of Results
  • Retrospective Studies
  • Visual Acuity / physiology