Development of a new biomechanically defined single impact rabbit cartilage trauma model for in vivo-studies

J Invest Surg. 2012 Aug;25(4):235-41. doi: 10.3109/08941939.2011.630123. Epub 2012 May 9.

Abstract

Background: Clinically oriented and easy to handle animal models are urgently needed to test pharmacologic treatment of cartilage trauma to reduce the resulting tissue damage by chondrocyte apoptosis and induction of matrix-degrading enzymes.

Aim: To develop a biomechanically defined cartilage trauma model.

Material and methods: We constructed a novel trauma device that allows biomechanically defined force application to the load-bearing region of the medial and lateral femoral condyles in adult rabbits. The fixation to the femur was specially designed to avoid uncontrolled influx of blood into the joint. The device was tested on the articular femoral surface of cadaveric rabbits.

Results: At a lower energy (1.0 J), the tests showed that superficial and partially deep fissuring, partial necrosis of the chondrocytes, and early proteoglycan loss occurred at the region of impact. Subchondral fractures could be excluded by micro CT. At higher energy (≥ 1.4 J), we observed more pronounced deep fissuring and in some cases complete shearing of the articular cartilage from the subchondral bone.

Conclusion: Our model represents an easy to use method to create a biomechanically defined cartilage trauma and offers some advantages with respect to handling under aseptic surgical conditions and prevention of uncontrolled intra-articular bleeding from the bone marrow compartment for pharmacologic studies.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Biomechanical Phenomena
  • Cartilage, Articular / injuries*
  • Cartilage, Articular / surgery
  • Femur / injuries
  • Rabbits