Cardiac output by Flotrac/VigileoTM validation trials: are there reliable conclusions?

Rev Recent Clin Trials. 2012 Aug;7(3):181-6. doi: 10.2174/157488712802281367.

Abstract

We reviewed the comparative trials of the Flotrac/VigileoTM versus the thermodilution method, published in the last five years. The results about the agreement between the two methods measuring cardiac output are contrasting. We also noticed that almost the whole pertinent literature include studies conducted without a correct statistical design, particularly about the sample size. For this reason we consider that results of the published studies about the agreement between pulse contour analysis for cardiac output measurement and thermodilution method may be not reliable.

Publication types

  • Review

MeSH terms

  • Cardiac Output / physiology*
  • Equipment Design
  • Heart Rate
  • Humans
  • Male
  • Monitoring, Intraoperative / instrumentation*
  • Randomized Controlled Trials as Topic*
  • Reproducibility of Results
  • Thermodilution / instrumentation