Level-set-based reconstruction algorithm for EIT lung images: first clinical results

Physiol Meas. 2012 May;33(5):739-50. doi: 10.1088/0967-3334/33/5/739. Epub 2012 Apr 24.

Abstract

We show the first clinical results using the level-set-based reconstruction algorithm for electrical impedance tomography (EIT) data. The level-set-based reconstruction method (LSRM) allows the reconstruction of non-smooth interfaces between image regions, which are typically smoothed by traditional voxel-based reconstruction methods (VBRMs). We develop a time difference formulation of the LSRM for 2D images. The proposed reconstruction method is applied to reconstruct clinical EIT data of a slow flow inflation pressure-volume manoeuvre in lung-healthy and adult lung-injury patients. Images from the LSRM and the VBRM are compared. The results show comparable reconstructed images, but with an improved ability to reconstruct sharp conductivity changes in the distribution of lung ventilation using the LSRM.

MeSH terms

  • Acute Lung Injury / diagnosis
  • Acute Lung Injury / therapy
  • Adult
  • Algorithms*
  • Electric Impedance
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Lung*
  • Middle Aged
  • Respiration, Artificial
  • Tomography / methods*