In vitro ultrasound characterization of a polyurethane trileaflet valve

ASAIO Trans. 1990 Jul-Sep;36(3):M532-5.

Abstract

Polyurethane synthetic trileaflet valves were compared with commercial prostheses in vitro, in a pulse duplicator using ultrasound to characterize the flow velocities and patterns. Flow-pressure drop behavior was in the middle range of other prosthetic valves. Diastolic regurgitant jets were located by color Doppler ultrasound, and there appeared to be some leakage through the leaflet fold at the commissure. Nevertheless, the closing volumes and closed valve leakage volumes were, on average, lower than other prosthetic valves. Systolic 20 Hz spectral oscillations detected with pulsed Doppler and continuous wave Doppler were attributed to leaflet flutter in the open valve.