Measurement of shear strength for HOPG with scanning tunneling microscopy by thermal excitation method

Ultramicroscopy. 2012 Apr:115:1-6. doi: 10.1016/j.ultramic.2011.01.018. Epub 2011 Jan 19.

Abstract

An experimental observation of force interactions in scanning tunneling microscopy (STM) is presented. A technique for measuring force interactions between a conventional STM probe and a sample by spectra analysis of its thermal fluctuations from tunneling current in STM is developed theoretically and experimentally. Thermally excited fluctuation of the STM probe is exactly discerned in air and then force gradient is determined from its corresponding eigen-frequency with a formula similar to that for a small-amplitude atomic force microscopy (AFM). The observed force interactions are consistent with forces in dynamic AFM. Shear strength of 7 GPa for highly oriented pyrolytic graphite (HOPG) under compressive stress is obtained from the experiment and using the elastic theory. We believe that this technique is of scientific significance as it enables accurate measurement of short-range force interactions at atomic scale under true STM conditions.

Publication types

  • Research Support, Non-U.S. Gov't